• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 HI-FLOW KIT; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ACE 68 HI-FLOW KIT; NRY Back to Search Results
Catalog Number 5MAXACE068KIT
Device Problems Unraveled Material (1664); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Patient Problem No Patient Involvement (2645)
Event Date 12/11/2017
Event Type  malfunction  
Manufacturer Narrative
Results: the penumbra system ace 68 reperfusion catheter (ace68) was fractured approximately 105.0 cm from the hub.Conclusions: evaluation of the returned device revealed that the ace68 was fractured.This damage may have occurred due to forceful handling of the ace68 at extreme angles during removal from the packaging.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
During preparation for a thrombectomy procedure, the hospital staff removed a penumbra system ace 68 reperfusion catheter (ace68) from the packaging of a penumbra system ace 68 high flow kit (kit) and went to straighten the ace68 out to flush it; however, the ace68 kept unraveling.The ace68 unraveled prior to use and therefore, was not used in the procedure.The procedure was completed using another kit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 HI-FLOW KIT
Type of Device
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7183916
MDR Text Key97754434
Report Number3005168196-2018-00118
Device Sequence Number1
Product Code NRY
UDI-Device Identifier00814548016603
UDI-Public00814548016603
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161640
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/17/2020
Device Catalogue Number5MAXACE068KIT
Device Lot NumberF77175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2017
Initial Date FDA Received01/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-