• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER TMT TRABECULAR METAL DISTAL FEMORAL AUGMENT BLOCK; TM DISTAL FEMORAL AUGMENT BLOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER TMT TRABECULAR METAL DISTAL FEMORAL AUGMENT BLOCK; TM DISTAL FEMORAL AUGMENT BLOCK Back to Search Results
Model Number N/A
Device Problems Device Slipped (1584); Metal Shedding Debris (1804); Naturally Worn (2988)
Patient Problems Pain (1994); Osteolysis (2377); Test Result (2695)
Event Date 11/01/2017
Event Type  Injury  
Event Description
It was reported that the patient underwent initial surgery on (b)(6) 2015.Almost 2 years later patient complained that his knee still had pain subsequently patient was revised on (b)(6) 2017 due to pain, wear, implant disassociation, loosening, metallosis, osteolysis.Tm augment is the only tm design controlled part.
 
Manufacturer Narrative
Patient is a (b)(6) male, medium build, no height, weight or activity level was reported.Patient has a history of 5 surgeries on his left knee spanning a 13 year period which includes the primary surgery in 2004 through the fourth revision surgery in 2017.Complaint alleges pain in the left knee, implant wear, implant disassociation, loosening, metallosis and osteolysis.The patient was revised on (b)(6) 2017.An engineering investigation was conducted by product development.The evidence while limited and not overwhelmingly conclusive, suggests that the suspected damage was primarily concentrated on the medial side of the femoral component where both a 5mm distal femoral augment block and a 5mm tm posterior femoral augment block were attached.Damage included a raised lip surrounding the internal thread on the femoral component where the distal augment screw could have pulled out from the internal threads as well as damage to the visible edge of the medial tm posterior femoral augment block.In addition, there appeared to be wear marks on the medial side of the femoral component where the 5mm distal femoral augment block was once attached.This implies that at least the nexgen rhk/lcck femoral augment screw which attached the 5mm distal femoral augment block loosened at some point in time which allowed relative motion between the two components.However, the independent radiologist who assessed the images reported that there was no evidence of wear or metallic particles seen on the films.Loosening of the nexgen rhk/lcck femoral augment screw was likely a result of failing to use the required augment torque wrench, as specified in the surgical technique and package insert, which ensures proper tightening.Failing to use the appropriate instrumentation was reported in a follow-up email from the sales rep.The 10mm distal femoral augment was also reported as being separated from the femoral augment and although not unreasonable to expect given that the same instrumentation was used, the limited evidence provided in this complaint does not support this.This investigation by product development is considered closed at this time.Should additional information become available to further this investigation, this report shall be updated.
 
Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRABECULAR METAL DISTAL FEMORAL AUGMENT BLOCK
Type of Device
TM DISTAL FEMORAL AUGMENT BLOCK
Manufacturer (Section D)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
MDR Report Key7185263
MDR Text Key97053950
Report Number3005751028-2018-00007
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
PK040487
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 05/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberN/A
Device Catalogue Number00549003620
Device Lot Number62684643
Other Device ID NumberN/A
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/05/2018
Initial Date FDA Received01/12/2018
Supplement Dates Manufacturer Received01/05/2018
05/01/2020
Supplement Dates FDA Received04/18/2018
05/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-