ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problem Break (1069)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
Manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported (b)(6) the patient experienced loss of stimulation.Xrays revealed the lead was broken.The patient underwent surgical intervention on (b)(6) 2017 where in the lead was explanted and replaced.Intraoperative diagnostics revealed high impedances on the extension.As a result the extension was replaced too (reference mfr report # 1627487-2018-00460) which resolved the issue.Stimulation was restored postoperatively.Patient was implanted sometime in 2006 (11 years ago) device information is unknown at this time.Additional devices may be added at a later date if needed based on available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7186195
MDR Text Key97102978
Report Number1627487-2018-00458
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3186
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-