Model Number 37800
Device Problem Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/09/2018
Event Type  Injury  
Manufacturer Narrative
The main component of the system and other applicable components are: product id: 4351-35, serial# (b)(4), lot# serial# (b)(4), implanted: (b)(6) 2013, explanted: (b)(6) 2018, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A manufacturing representative (rep) reported on behalf of a healthcare provider (hcp) that the patient was in a revision and they replaced the implantable neurostimulator.They tested impedance and it was out of range.They replaced the leads and the impedance was fine.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7187694
MDR Text Key97270355
Report Number3004209178-2018-00749
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2019
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/09/2018
Date Device Manufactured07/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age37 YR
-
-