• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G141
Device Problems Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Atrial Tachycardia (1731)
Event Date 01/17/2017
Event Type  Injury  
Manufacturer Narrative
Additional information received from the local field representative indicated that the device sensitivity was reprogrammed to 1.0 mv and the oversensing resolved.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) lead were oversensing atrial tachycardia that resulted in asystole for approximately 6 seconds.The issue was also observed with the previous cardiac resynchronization therapy defibrillator (crt-d).Boston scientific technical services (ts) recommended programming the sensitivity from nominal at 0.6 mv to 1.5 mv and performing defibrillation threshold (dft) testing to ensure appropriate ventricular fibrillation (vf) detection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7190201
MDR Text Key97281978
Report Number2124215-2017-19892
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534553
UDI-Public00802526534553
Combination Product (y/n)N
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/14/2018
Device Model NumberG141
Other Device ID NumberINOGEN CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received10/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
-
-