Boston scientific received information that during an upgrade procedure when another manufacturer's chronic right atrial (ra) lead was connected to this cardiac resynchronization therapy pacemaker (crt-p) high out of range pacing impedance measurements of greater than 3000 ohms was noted.The physician disconnected the lead and reconnected with the same result.The lead was disconnected again and the port was flushed then the lead was reconnected and still exhibited greater than 3000 ohms in both bipolar and unipolar configurations.The physician disconnected and reconnected the lead a third time with good impedance measurements.However, the physician expressed concern over a possible issue with the crt-p.The physician commented that he may have seen something in the ra port, but the field representative noted he did not see anything.A new crt-p was successfully connected and gave good measurements with all leads.This crt-p was attempted and not implanted.The chronic ra lead was left implanted with the new device.No adverse patient effects were reported.
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header, lead ports and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
|