Model Number N/A
Device Problem Expulsion (2933)
Patient Problem No Patient Involvement (2645)
Event Date 04/10/2017
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Investigation summary: review of the device history record for (b)(4), lot number z000005514, identified no relevant deviations or anomalies.Product examination could not be performed as no product was returned for this complaint.However, the customer provided photos that show the battery pack had ruptured.The photos indicated that the unit had not been used.This complaint is confirmed.The reported event claimed that the sealed unit had ruptured before it had been opened.While rare, this kind of event occurs when the batteries overheat from a short circuit.To address this issue, a change notice has been implemented to adjust the length of the wires inside the battery pack.This means that there will no longer be a need to tightly fold wires inside the battery pack in order for them to reach their respective circuit contacts.This reduces the risk of a short circuit.The root cause of the reported event could not be specifically determined with the provided information.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Device not returned.
 
Event Description
It was reported that the battery pack of the pulse lavage burst in the package before it had been opened and used.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key7197920
MDR Text Key97882208
Report Number0001526350-2018-00064
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model NumberN/A
Device Catalogue Number00515047500
Device Lot NumberZ000005514
Other Device ID Number(01)00889024375154
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/16/2018
Initial Date FDA Received01/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-