LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Electrical /Electronic Property Problem (1198); Loss of Power (1475)
Patient Problem No Patient Involvement (2645)
Event Date 12/19/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective compressor.The defective heater-cooler was requested back to livanova (b)(4) for repair and investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t shuts down on its own and causes a short circuit in the operation room.The issue has been identified during maintenance.There was no patient involvement.
 
Manufacturer Narrative
The defective device was returned for further investigation.During the investigation the reported failure could be confirmed.As root cause a hole inside the evaporator could be identified.Corrective actions are in progress for this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7198638
MDR Text Key97879131
Report Number9611109-2018-00044
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/19/2017
Initial Date FDA Received01/17/2018
Supplement Dates Manufacturer Received02/23/2018
Supplement Dates FDA Received03/23/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-