Model Number E100-29M
Device Problem Torn Material (3024)
Patient Problem Aortic Insufficiency (1715)
Event Date 11/24/2017
Event Type  Injury  
Event Description
On (b)(6) 2014, a mitral valve replacement was performed and this 29 mm epic tissue valve was implanted.On (b)(6) 2017, the valve was explanted due to a rupture observed in one of the cusps.A 31 mm biocor pericardial heart valve (model: b30-31m; sn: (b)(4)) was implanted.The patient was reported to be in stable condition.
 
Manufacturer Narrative
Product investigation: the reported event of a "rupture observed in one of the cusps" was confirmed.Gross morphological and histopathological examination found cusps 1 and 2 were torn.Cusp 2 appeared to be folded in on itself.No acute inflammation or significant calcifications were found.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA.
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR  31310-260
Manufacturer (Section G)
ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR   31310-260
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7199274
MDR Text Key97507098
Report Number3001883144-2018-00014
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/10/2017
Device Model NumberE100-29M
Device Catalogue NumberE100-29M
Other Device ID Number05414734027533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age51 YR
Patient Weight70
-
-