Event summary: upon visual inspection of the flexcath sheath (4fc12 / 92475-094) results showed the device was not intact and was kinked on the 2.5460 and 1.1315 inches from the tip of the sheath.The sheath failed the test due to a kink near the tip of the sheath.In conclusion, the reported sheath kink issue has been confirmed through testing.The sheath failed the inspection due to a kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the sheath was broken.The sheath was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.(b)(6) 2018: the sheath was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|