Model Number N/A |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/04/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that oxford femoral drill was broken at total knee arthroplasty upon drilling subsequently alternative drill was used to complete the procedure.
|
|
Manufacturer Narrative
|
(b)(4).Products have been returned to (b)(4) for evaluation and forwarded to the complaints and vigilance engineer.A complaint history review was conducted for (b)(4).The search identified 6 further complaints for the same type of issue for the same lot # zb120701 between the dates of (b)(6) 2015 to (b)(6) 2018.The complaints data will be presented to the capa review board due to a trend being found as per local procedure sop-163 rev 6 - data trend analysis.A visual inspection of the instrument showed evidence of heavy wear which suggests the most likely cause of the fracture to be wear and tear.The review of the documentation has confirmed that the instrument was conforming to pre-defined drawing specification when manufactured in 2013.
|
|
Search Alerts/Recalls
|