MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1712K
Device Problems Component Missing (2306); Material Deformation (2976); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/15/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.  however, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that insulin pump had cosmetic damage.Customer¿s blood glucose was unknown.Customer reported that ring on top of reservoir compartment is missing and reservoir cannot lock into place anymore.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Unable to perform displacement test due to missing retainer.The reservoir does not stay locked in due to missing retainer.
 
Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0958-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1712K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7215761
MDR Text Key98167281
Report Number3004209178-2018-45776
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169739512
UDI-Public(01)00643169739512
Combination Product (y/n)N
Reporter Country CodeBE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712K
Device Catalogue NumberMMT-1712K
Device Lot NumberHG1UBNZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/15/2018
Initial Date FDA Received01/24/2018
Supplement Dates Manufacturer Received08/09/2018
10/14/2019
Supplement Dates FDA Received08/31/2018
05/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-0958-2020
Patient Sequence Number1
-
-