Model Number G158 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
Atrial Flutter (1730)
|
Event Date 12/19/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the device remains in service.If additional information is received, this report will be updated.
|
|
Event Description
|
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) exhibited right ventricular (rv) oversensing of right atrial (ra) signals, which resulted in greater than two seconds of pacing inhibition.Additionally, the patient¿s atrial flutter was observed on the rv and shock channels; however was not being oversensed by the device.The device was reprogrammed to resolve the oversensing.No additional adverse patient effects were reported.The device remains in service.
|
|
Event Description
|
Additional information was received which noted that a surgical revision was performed.The patient¿s rv lead was explanted and replaced, and this device remains in service.No additional adverse patient effects were reported.
|
|