No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.It was reported that the computer for the navigation system was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The computer for the navigation system was returned to the manufacturer for evaluation.Testing found that there were 3 bad sectors on the hard drive.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
A site representative reported that, while outside of a procedure, the navigation system displayed a sr manager failure.After multiple restarts, the functionality to the navigation system was restored.There was no patient present when this issue was identified.No additional information was provided.
|