MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87216
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Device Handling Problem (3265)
Patient Problem Peritonitis (2252)
Event Date 12/22/2017
Event Type  Injury  
Manufacturer Narrative
The device was not returned to the manufacturer for physical evaluation, and the lot number was not able to be obtained to date.Distribution records were reviewed to identify potential lots of this product shipped to the customer over the past 3 months.The entire set of lots have been sold and distributed.Batch records for the lots identified were reviewed and confirmed there were no deviations or non-conformances during the manufacturing process.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.There is no documentation in the complaint file supporting an association between the liberty cycler or the liberty cycler set and the event of peritonitis.Additionally, there is no allegation for this time period against any fresenius devices in relation to this event.However, in light of the pd nurse reporting the adverse event was attributed to touch contamination; a strong likely relationship with breaks in aseptic technique during pd treatments and the resulting episode of peritonitis exists.
 
Event Description
A peritoneal dialysis patient's nurse reported the patient utilizing continuous cyclic peritoneal dialysis [cc(pd)] therapy for renal replacement therapy (rrt) presented to the pd clinic on (b)(6) 2017 with a cloudy effluent bag.The patient was diagnosed with peritonitis with reported growth of coagulase negative staphylococcus.According to the patients nurse the event was attributed to a breach in technique (details unknown).The patient was not hospitalized and was prescribed antibiotics levofloxacin and vancomycin (route, dose and duration unknown).Accordingly, the patient 's peritonitis has cleared with white blood cells (wbc) reducing (level unknown) in subsequent effluent checks.The patient was not hospitalized for the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE CONN./EXT. DL
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key7217543
MDR Text Key98162041
Report Number8030665-2018-00131
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861100750
UDI-Public00840861100750
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue Number050-87216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Initial Date Manufacturer Received 01/05/2018
Initial Date FDA Received01/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient Weight74
-
-