During preparation for a thrombectomy procedure, the hospital staff found that the penumbra system ace 68 reperfusion catheter (ace68) was broken upon removal from the packaging.The damage to the ace68 was found prior to use, and therefore it was not used in the procedure.The procedure was completed using a new ace68.
|
Results: the ace68 was fractured approximately 39.0 cm from the hub, ovalized approximately 110.0 and 130.0 cm from the hub, and kinked approximately 125.0 cm from the hub.Conclusions: evaluation of the returned device revealed the ace68 was kinked, ovalized, and fractured.This damage may have occurred due to forceful handling of the ace68 during removal from the packaging or preparation for use.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|