Model Number FUSION |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Display or Visual Feedback Problem (1184); Device Displays Incorrect Message (2591); Environmental Compatibility Problem (2929)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.A medtronic representative went to site to test the equipment.Representative was able to replicate the reported issue as the system displayed "no input detected" message and fan cycling sound could be heard.Rebooting the system did not resolve the issue.Representative reported that a computer was replaced to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect computer has been received by the manufacturer for evaluation.The computer was tested and the issue was confirmed as the computer was power cycling.
|
|
Event Description
|
A medtronic representative reported that during functional endoscopic sinus surgery (fess),the navigation system unexpectedly shut down and booted up with "no input detected".Rebooting the system resolved the issue temporarily and site proceeded with surgery.The procedure was completed with the use of navigation.No information was provided on delay to procedure.No known impact on patient outcome.
|
|
Manufacturer Narrative
|
The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|
Manufacturer Narrative
|
Additional information: patient demographics provided.The surgeon opted to complete the procedure without the use of the navigation system.There was no reported impact on patient outcome.Correction: event date updated to proper value.
|
|
Manufacturer Narrative
|
The fda codes were updated to reflect current coding guidelines.If information is provided in the future, a supplemental report will be issued.
|
|