• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX50012UX
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 12/01/2017
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
During the procedure a resolute onyx rx drug eluting stent was deployed in the rca, at 12 atm.It was reported that the stent embolized into the guide catheter.The guide was retracted however, the stent remained in the right subclavian.The stent was snared and pulled back into the guide catheter and removed from the body intact.There was no patient injury reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7230397
MDR Text Key98669745
Report Number9612164-2018-00179
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169557291
UDI-Public00643169557291
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Catalogue NumberRONYX50012UX
Device Lot Number0008735235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/29/2018
Date Device Manufactured07/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age93 YR
-
-