ROCHE DIAGNOSTICS ELECSYS FT4 II ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number ASKU
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2018
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer stated that they received erroneous results for one patient sample tested for elecsys ft3 iii (ft3) and the elecsys ft4 ii assay (ft4) on a cobas 6000 e 601 module (e601).The erroneous results were reported outside of the laboratory to medical personnel treating the patient.The patient was tested after having a thyroid operation and the doctor expected results indicating hypothyroidism.(b)(4).Refer to the attachment for all patient data.The sample was initially tested on the e601 analyzer on (b)(6) 2018 and was later repeated on a beckman analyzer at a different site on (b)(6) 2018.No adverse events were alleged to have occurred with the patient.The e601 analyzer serial number was (b)(4).
 
Manufacturer Narrative
A sample from the patient was provided for investigation.The ft4 value obtained by the customer could not be duplicated.The ft4 value was found to be below the normal reference range of the assay.The investigation was unable to find a definitive root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 II ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key7230882
MDR Text Key98986648
Report Number1823260-2018-00286
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot Number270341
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age77 YR
-
-