• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG ILED DUO; ILED 3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG ILED DUO; ILED 3 Back to Search Results
Model Number 1565160
Device Problems Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/05/2018
Event Type  malfunction  
Manufacturer Narrative
The components with chipping paint will be replaced.Our investigation has identified steps in the manufacturing process which combined with non-approved cleaners can lead to the paint chipping.The damage to the painted surfaces usually does not develop suddenly, but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The surface coating process will be updated to include additional testing and verification.
 
Event Description
Trumpf medical was informed by a customer that paint from the surface of their surgical lights was chipping.No paint fell into the sterile field and no injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILED DUO
Type of Device
ILED 3
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key7235272
MDR Text Key99001499
Report Number9681407-2018-00009
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Remedial Action Replace
Type of Report Initial
Report Date 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1565160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-