BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Analysis revealed no damage to the capio slim suture capturing device.The needle dart was retrieved from behind the cage.There is no amount of suture attached to the needle indicating that the needle pulled off of the suture.The uphold mesh assembly was not returned.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Review and analysis of all available information fails to indicate a probable cause.Therefore, the assigned complaint investigation conclusion code for this event is undeterminable.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a cystocele repair procedure on (b)(6) 2017.According to the complainant, during the procedure, the needle did not penetrate the tissue when the capio slim was actuated to implant the mesh leg on the right side.The procedure was completed with another uphold lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.This event has been deemed reportable based on the investigation results: the needle dart was retrieved from behind the cage of the suturing device.There is no amount of suture attached to the needle indicating that the needle pulled off of the suture.
|
|
Search Alerts/Recalls
|
|
|