During the procedure, an endostitch was being used.At one point, the stitch became stuck in the device.While trying to retrieve it, the needle broke in half.The device was removed from the field and replaced with a new device.The pieces of the needle were retrieved with no harm to the patient and the procedure was completed as planned.Upon receipt of the device, clinical engineering also noted the black button on the handle of the device was broken.It is unclear whether this was the cause of, or the result of, the device becoming jammed.Images are included in the attachments section of this report.A return kit will be issued and the device will be returned for failure analysis.
|