Results: the ace68 was kinked approximately 59.0 cm from the proximal end.Conclusions: evaluation of the returned device confirmed the ace68 was kinked.If the ace68 is withdrawn from the packaging prior to removing the tubing tray from the kit retainer, damage such as this may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|