• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM07080
Device Problems Positioning Failure (1158); Fracture (1260); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/08/2018
Event Type  malfunction  
Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer; however a photo was provided and a photo review will be performed.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation to date.The investigation of the reported event is currently underway the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that during a stent graft placement procedure for the cephalic arch, the device was allegedly difficult to advance to the target lesion.It was further reported that the outer sheath of the device allegedly fractured during an attempt to deploy the stent graft.Reportedly, the device was removed from the patient without issue and two new stent grafts were placed using the overlapping technique.There was no reported patient injury.
 
Event Description
It was reported that during a stent graft placement procedure for the cephalic arch, the device was allegedly difficult to advance to the target lesion.It was further reported that the outer sheath of the device allegedly fractured during an attempt to deploy the stent graft.Reportedly, the device was removed from the patient without issue and two new stent grafts were placed using the overlapping technique.There was no reported patient injury.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was performed.The lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.No additional complaint has been reported for this lot number previously.Investigation summary: based on the condition of the returned sample and the photo provided, the reported issue could be confirmed.The stent graft was found partially deployed and the outer sheath was found elongated which indicates that increased friction affected the delivery system during attempt of stent graft deployment and finally led to the fracture of the outer sheath.This fracture makes a stent graft deployment impossible.Potential factors which may have caused or contributed to the reported issue have been considered.Based on the information available and the evaluation of the sample returned, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risks.Regarding preparation of the device the ifu states that "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure" furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." regarding the use of accessories the ifu states: "materials required for the fluency® plus endovascular stent graft procedure: (.) introducer sheath with appropriate inner diameter (.)".(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7241212
MDR Text Key99008928
Report Number2020394-2018-00052
Device Sequence Number1
Product Code PFV
UDI-Device Identifier04049519008455
UDI-Public(01)04049519008455
Combination Product (y/n)N
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/05/2020
Device Model NumberFEM07080
Device Catalogue NumberFEM07080
Device Lot NumberANBS0470
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Date Manufacturer Received03/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Weight91
-
-