• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION Back to Search Results
Model Number 3341
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/15/2017
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 3.Reference mfr.Report#: 1627487-2018-0957.Reference mfr.Report#: 3006705815-2018-00211.It was reported the patient stated while reaching for something on (b)(6) 2017, she heard a "pop" in her shoulder.The next day, the patient had lost stimulation.As such, a company representative met with the patient for reprogramming which initially resolved the issue.However, the issue reoccurred.Diagnostic testing revealed high impedance values on multiple lead contacts.Imaging revealed the extensions had pulled out of the port 1-8 and all lead contacts have high impedance values.As a result, the patient underwent surgical intervention on (b)(6) 2018 where the ipg was explanted and replaced.In addition, one extension was explanted and 3 were implanted.Please note, it is unknown which extension was explanted.Therefore, all suspected devices are being reported.
 
Event Description
Device 1 of 5- reference mfr.Report#: 1627487-2018-00957; reference mfr.Report#: 3006705815-2018-00211; reference mfr.Report#: 1627487-2018-03315; reference mfr.Report#: 1627487-2018-03316.Please note: two model 3383 extensions have been added to this report as ( devices 4 and 5).The analysis of the extensions revealed the device tested out of specification in a manner that relates to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7242064
MDR Text Key99040876
Report Number1627487-2018-00955
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model Number3341
Device Lot Number5459843
Other Device ID Number05414734402279
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/10/2018
Initial Date FDA Received02/05/2018
Supplement Dates Manufacturer Received03/07/2017
Supplement Dates FDA Received04/02/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192(40, SCS ANCHOR; MODEL 3163 (4), SCS LEAD; MODEL 3383 (2), SCS EXTENSION
Patient Outcome(s) Other;
Patient Age48 YR
-
-