BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 01/11/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold (tm) lite w/ capio slim device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018 according to the complainant, during the procedure, the needle detached on the left side of the patient while implanting the first leg.Reportedly, it was too small to be removed and was left inside the patient.The procedure was completed with another uphold¿ lite w/ capio slim.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
A visual examination of the returned uphold lite and capio slim revealed that the suture on the blue dilator is broken.The remainder of the suture with dart was not returned.There is no damage to the suture capturing device.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Analysis reveals that the suture is severed, which could be due to interaction with the sharp edge of the carrier.Therefore, the assigned complaint investigation conclusion code for this event is manufacturing process design because the design or validation of the manufacturing process was not sufficient to ensure the finished device met the intent of the design.The issue is under investigation and a correction has not yet been implemented.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold (tm) lite with capio slim device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the needle detached on the left side of the patient while implanting the first leg.Reportedly, it was too small to be removed and was left inside the patient.The procedure was completed with another uphold lite with capio slim.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|