• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 01/11/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold (tm) lite w/ capio slim device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018 according to the complainant, during the procedure, the needle detached on the left side of the patient while implanting the first leg.Reportedly, it was too small to be removed and was left inside the patient.The procedure was completed with another uphold¿ lite w/ capio slim.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual examination of the returned uphold lite and capio slim revealed that the suture on the blue dilator is broken.The remainder of the suture with dart was not returned.There is no damage to the suture capturing device.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Analysis reveals that the suture is severed, which could be due to interaction with the sharp edge of the carrier.Therefore, the assigned complaint investigation conclusion code for this event is manufacturing process design because the design or validation of the manufacturing process was not sufficient to ensure the finished device met the intent of the design.The issue is under investigation and a correction has not yet been implemented.
 
Event Description
It was reported to boston scientific corporation that an uphold (tm) lite with capio slim device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the needle detached on the left side of the patient while implanting the first leg.Reportedly, it was too small to be removed and was left inside the patient.The procedure was completed with another uphold lite with capio slim.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7243297
MDR Text Key99069846
Report Number3005099803-2018-00171
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/12/2020
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000058945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
Patient Weight51
-
-