• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. GUIDE WIRE GRIPPER; PASSER, WIRE, ORTHOPEDIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. GUIDE WIRE GRIPPER; PASSER, WIRE, ORTHOPEDIC Back to Search Results
Model Number N/A
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/07/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign.The event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported the wire gripper was no longer gripping the wire.The surgery was completed with another device without delay.No further information is available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed impact marks and scratches on the impaction surface.Both holes exhibit heavy gauging and wear indicative of use.Also noted impaction marks on the product.The wear on the holes leads to insufficient gripping of guide wires.Device history record (dhr) was reviewed and no discrepancies were found relevant to the reported event.Investigation results concluded the root cause of the event is normal wear during the instrument's field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE WIRE GRIPPER
Type of Device
PASSER, WIRE, ORTHOPEDIC
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7244487
MDR Text Key99126314
Report Number0001822565-2018-00648
Device Sequence Number1
Product Code HXI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00249001200
Device Lot Number63107373
Other Device ID Number(01) 00889024063136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-