EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 TRANSCATHETER HEATH VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9600TFX
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Injury (2348)
Event Date 11/15/2017
Event Type  Injury  
Manufacturer Narrative
Exemption number e2016006, number of serious injury events 1763.Column a indicates whether an event is a ¿new event or follow-up¿.Columns m and n associate a device with the event type (e.G.It is considered most appropriate to assign an event of ¿major vascular complication¿ to a sheath, rather than the sapien 3 valve).Associated device information is not included in the data received from the registry; the products were assigned by edwards based on standard kitting and the ¿implant approach¿ field available in the registry.Any line items highlighted in green indicate use of a sapien 3 valve with either a transapical or transaortic approach prior to the commercial approval of the certitude delivery system.Although the associated device is listed as the ¿edwards certitude introducer sheath set¿, it is possible that the ¿ascendra+ introducer sheath set¿ was the actual sheath used.This specific detail is not provided to edwards.Any line items highlighted in yellow indicate a discrepancy in the registry data; example: ¿event date¿ prior to ¿implant date¿.
 
Event Description
Thv/tvt registry alternative summary report (asr) adverse event submission for november 2017 data extract for serious injuries for the sapien 3 valve.
 
Event Description
Unique Complaint ID Number,Initial or Supplement,Type of event,TTE in days,Device Brand Name,Medical device identifier,Device Codes
3929982,I,SI,293,Edwards SAPIEN 3,9600TFX20,3190
3938481,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
3944682,I,SI,12,Edwards SAPIEN 3,9600TFX26,3190
3954942,I,SI,43,Edwards SAPIEN 3,9600TFX26,3190
3954942,I,SI,116,Edwards SAPIEN 3,9600TFX26,3190
3962025,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
3971147,I,SI,333,Edwards SAPIEN 3,9600TFX23,3190
3974415,I,SI,298,Edwards SAPIEN 3,9600TFX26,2993
3983941,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4011455,I,SI,303,Edwards SAPIEN 3,9600TFX29,3190
3971768,I,SI,380,Edwards SAPIEN 3,9600TFX26,3190
4029312,I,SI,130,Edwards SAPIEN 3,9600TFX26,3190
4029312,I,SI,228,Edwards SAPIEN 3,9600TFX26,3190
4055292,I,SI,120,Edwards SAPIEN 3,9600TFX29,3190
4071709,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4183004,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4186481,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
3810793,I,SI,229,Edwards SAPIEN 3,9600TFX26,3190
4244845,I,SI,19,Edwards SAPIEN 3,9600TFX26,2993
4257771,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4263063,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4282304,I,SI,51,Edwards SAPIEN 3,9600TFX23,3190
4289880,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4294992,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4296328,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4317206,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4319090,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4322206,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4329658,I,SI,34,Edwards SAPIEN 3,9600TFX29,3190
4343194,I,SI,31,Edwards SAPIEN 3,9600TFX23,2993
4366196,I,SI,-40,Edwards SAPIEN 3,9600TFX26,3190
4366293,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4369428,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
1132048,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
1579342,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
1579342,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
2102290,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
2102290,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
2151,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
2172413,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
2478468,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
2574874,I,SI,-1585,Edwards SAPIEN 3,9600TFX26,2993
2748831,I,SI,35,Edwards SAPIEN 3,9600TFX23,3190
2835797,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
2850189,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
2884951,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
2884951,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
2932526,I,SI,10,Edwards SAPIEN 3,9600TFX20,3190
2933439,I,SI,15,Edwards SAPIEN 3,9600TFX29,2993
2978983,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
3083134,I,SI,40,Edwards SAPIEN 3,9600TFX29,3190
3131528,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
3292491,I,SI,259,Edwards SAPIEN 3,9600TFX26,3190
3422312,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
3422312,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
3424865,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
3670837,I,SI,-353,Edwards SAPIEN 3,9600TFX23,3190
3689097,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
3813125,I,SI,315,Edwards SAPIEN 3,9600TFX23,3190
3821221,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
3844519,I,SI,109,Edwards SAPIEN 3,9600TFX29,3190
3876037,I,SI,162,Edwards SAPIEN 3,9600TFX26,3190
3914365,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
3921848,I,SI,267,Edwards SAPIEN 3,9600TFX26,3190
3928127,I,SI,127,Edwards SAPIEN 3,9600TFX23,3190
3928127,I,SI,410,Edwards SAPIEN 3,9600TFX23,3190
3932250,I,SI,329,Edwards SAPIEN 3,9600TFX23,3190
3932250,I,SI,329,Edwards SAPIEN 3,9600TFX23,3190
3942404,I,SI,175,Edwards SAPIEN 3,9600TFX26,3190
3942404,I,SI,175,Edwards SAPIEN 3,9600TFX26,3190
3951614,I,SI,265,Edwards SAPIEN 3,9600TFX29,3190
3953301,I,SI,29,Edwards SAPIEN 3,9600TFX23,3190
3955970,I,SI,17,Edwards SAPIEN 3,9600TFX26,2993
3963282,I,SI,340,Edwards SAPIEN 3,9600TFX29,3190
3965372,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
3970037,I,SI,124,Edwards SAPIEN 3,9600TFX23,3190
3970037,I,SI,162,Edwards SAPIEN 3,9600TFX23,3190
3970037,I,SI,437,Edwards SAPIEN 3,9600TFX23,3190
3971054,I,SI,214,Edwards SAPIEN 3,9600TFX26,3190
3971321,I,SI,144,Edwards SAPIEN 3,9600TFX26,3190
3974495,I,SI,410,Edwards SAPIEN 3,9600TFX26,3190
3978450,I,SI,365,Edwards SAPIEN 3,9600TFX23,3190
3979362,I,SI,6,Edwards SAPIEN 3,9600TFX29,3190
3982451,I,SI,199,Edwards SAPIEN 3,9600TFX23,3190
3997634,I,SI,96,Edwards SAPIEN 3,9600TFX29,3190
3997634,I,SI,156,Edwards SAPIEN 3,9600TFX29,3190
4000160,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4003642,I,SI,300,Edwards SAPIEN 3,9600TFX23,3190
4013095,I,SI,356,Edwards SAPIEN 3,9600TFX29,3190
4016843,I,SI,367,Edwards SAPIEN 3,9600TFX23,3190
4025666,I,SI,96,Edwards SAPIEN 3,9600TFX29,3190
4029118,I,SI,259,Edwards SAPIEN 3,9600TFX23,3190
4030325,I,SI,239,Edwards SAPIEN 3,9600TFX23,3190
4045501,I,SI,152,Edwards SAPIEN 3,9600TFX23,3190
4045864,I,SI,253,Edwards SAPIEN 3,9600TFX29,3190
4049906,I,SI,264,Edwards SAPIEN 3,9600TFX26,3190
4049906,I,SI,314,Edwards SAPIEN 3,9600TFX26,3190
4053840,I,SI,76,Edwards SAPIEN 3,9600TFX26,3190
4053840,I,SI,112,Edwards SAPIEN 3,9600TFX26,3190
4077293,I,SI,289,Edwards SAPIEN 3,9600TFX23,3190
4077298,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
4079523,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4084500,I,SI,8,Edwards SAPIEN 3,9600TFX29,2993
4090276,I,SI,189,Edwards SAPIEN 3,9600TFX26,3190
4091179,I,SI,329,Edwards SAPIEN 3,9600TFX23,2993
4091179,I,SI,320,Edwards SAPIEN 3,9600TFX23,2993
4096783,I,SI,119,Edwards SAPIEN 3,9600TFX26,3190
4127078,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4129171,I,SI,30,Edwards SAPIEN 3,9600TFX26,3190
4130847,I,SI,212,Edwards SAPIEN 3,9600TFX29,3190
4130847,I,SI,210,Edwards SAPIEN 3,9600TFX29,3190
4131189,I,SI,253,Edwards SAPIEN 3,9600TFX29,3190
4134493,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4143810,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4143940,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4157970,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4165830,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4165830,I,SI,4,Edwards SAPIEN 3,9600TFX26,3190
4168010,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4168734,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4168734,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4168757,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4175079,I,SI,1,Edwards SAPIEN 3,9600TFX29,3190
4175079,I,SI,1,Edwards SAPIEN 3,9600TFX29,3190
4192597,I,SI,38,Edwards SAPIEN 3,9600TFX20,3190
4194597,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4194597,I,SI,8,Edwards SAPIEN 3,9600TFX23,3190
4194635,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4194635,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4194652,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4197553,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4198836,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4199917,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4204439,I,SI,25,Edwards SAPIEN 3,9600TFX23,2993
4215239,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
4216266,I,SI,30,Edwards SAPIEN 3,9600TFX23,2993
4229455,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4231653,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4231717,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4240182,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4240972,I,SI,27,Edwards SAPIEN 3,9600TFX29,3190
4244484,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4244484,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4244484,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4244484,I,SI,10,Edwards SAPIEN 3,9600TFX23,3190
4250647,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4251857,I,SI,336,Edwards SAPIEN 3,9600TFX26,3190
4252167,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4252486,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4261080,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4262270,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4264097,I,SI,19,Edwards SAPIEN 3,9600TFX23,3190
4265156,I,SI,21,Edwards SAPIEN 3,9600TFX26,2993
4266005,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4266005,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4273949,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4276686,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4277229,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4277229,I,SI,31,Edwards SAPIEN 3,9600TFX29,3190
4277229,I,SI,36,Edwards SAPIEN 3,9600TFX29,3190
4278395,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4278400,I,SI,10,Edwards SAPIEN 3,9600TFX20,2993
4278469,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4278469,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4278469,I,SI,9,Edwards SAPIEN 3,9600TFX29,3190
4278655,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4280009,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4282694,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4284367,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4284367,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4284462,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4285250,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4285350,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4286083,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4288578,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4290167,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4290934,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4294245,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4297035,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4297298,I,SI,6,Edwards SAPIEN 3,9600TFX23,3190
4297298,I,SI,6,Edwards SAPIEN 3,9600TFX23,3190
4298577,I,SI,19,Edwards SAPIEN 3,9600TFX23,2993
4299791,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4300332,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4301163,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4305453,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4305453,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4305453,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4305886,I,SI,7,Edwards SAPIEN 3,9600TFX23,2993
4307322,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4307322,I,SI,5,Edwards SAPIEN 3,9600TFX23,3190
4309245,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4312786,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4312786,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4312789,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4313588,I,SI,97,Edwards SAPIEN 3,9600TFX26,3190
4315416,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4315426,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4317933,I,SI,22,Edwards SAPIEN 3,9600TFX29,2993
4319221,I,SI,44,Edwards SAPIEN 3,9600TFX23,3190
4319358,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4319358,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4319358,I,SI,6,Edwards SAPIEN 3,9600TFX23,3190
4320333,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4320333,I,SI,27,Edwards SAPIEN 3,9600TFX29,2993
4320333,I,SI,5,Edwards SAPIEN 3,9600TFX29,3190
4320492,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4322820,I,SI,17,Edwards SAPIEN 3,9600TFX29,3190
4322820,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4322820,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4323852,I,SI,12,Edwards SAPIEN 3,9600TFX23,2993
4324922,I,SI,42,Edwards SAPIEN 3,9600TFX26,3190
4325094,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4326335,I,SI,27,Edwards SAPIEN 3,9600TFX26,2993
4328410,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4328413,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4328413,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4328413,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4329406,I,SI,55,Edwards SAPIEN 3,9600TFX23,3190
4329406,I,SI,55,Edwards SAPIEN 3,9600TFX23,3190
4329682,I,SI,2,Edwards SAPIEN 3,9600TFX20,2993
4329972,I,SI,13,Edwards SAPIEN 3,9600TFX23,2993
4330950,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4331151,I,SI,7,Edwards SAPIEN 3,9600TFX23,2993
4335335,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4335632,I,SI,28,Edwards SAPIEN 3,9600TFX26,2993
4336002,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4336313,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4336332,I,SI,12,Edwards SAPIEN 3,9600TFX29,2993
4337581,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4338252,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4340230,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4340230,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4340662,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4340662,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4341469,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4341469,I,SI,4,Edwards SAPIEN 3,9600TFX26,3190
4343223,I,SI,10,Edwards SAPIEN 3,9600TFX23,3190
4343245,I,SI,49,Edwards SAPIEN 3,9600TFX29,3190
4344579,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4345204,I,SI,6,Edwards SAPIEN 3,9600TFX29,3190
4346515,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4346515,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4347825,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4348545,I,SI,21,Edwards SAPIEN 3,9600TFX29,2993
4350209,I,SI,20,Edwards SAPIEN 3,9600TFX23,3190
4350209,I,SI,60,Edwards SAPIEN 3,9600TFX23,3190
4350209,I,SI,40,Edwards SAPIEN 3,9600TFX23,3190
4350700,I,SI,62,Edwards SAPIEN 3,9600TFX29,3190
4351734,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4352319,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4353493,I,SI,22,Edwards SAPIEN 3,9600TFX23,2993
4353523,I,SI,77,Edwards SAPIEN 3,9600TFX23,3190
4353862,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4353865,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4353865,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4354951,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4357481,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4358323,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4358323,I,SI,1,Edwards SAPIEN 3,9600TFX29,3190
4358323,I,SI,1,Edwards SAPIEN 3,9600TFX29,3190
4359661,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4359661,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4359661,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4359661,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4359661,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4359667,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4360914,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4363618,I,SI,3,Edwards SAPIEN 3,9600TFX23,3190
4364095,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4364095,I,SI,7,Edwards SAPIEN 3,9600TFX26,3190
4364980,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4366197,I,SI,25,Edwards SAPIEN 3,9600TFX26,3190
4368367,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4369402,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4369402,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4369402,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4369402,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4369402,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4369441,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4369528,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4369693,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4369706,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4369726,I,SI,9,Edwards SAPIEN 3,9600TFX29,2993
4369929,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4370822,I,SI,28,Edwards SAPIEN 3,9600TFX26,2993
4371254,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4372097,I,SI,22,Edwards SAPIEN 3,9600TFX23,2993
4372745,I,SI,15,Edwards SAPIEN 3,9600TFX23,2993
4373417,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4373417,I,SI,8,Edwards SAPIEN 3,9600TFX23,3190
4374312,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4376251,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4377504,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4377598,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4377767,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4377847,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4378106,I,SI,27,Edwards SAPIEN 3,9600TFX29,3190
4378152,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4378161,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4378161,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4378164,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4378188,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4378481,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4378511,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4378578,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4378623,I,SI,13,Edwards SAPIEN 3,9600TFX26,2993
4378623,I,SI,13,Edwards SAPIEN 3,9600TFX26,3190
4378637,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4378697,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4378733,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4378857,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4379032,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4379244,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4379354,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4379366,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4379572,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4379637,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4379780,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4380361,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4380613,I,SI,12,Edwards SAPIEN 3,9600TFX29,3190
4380673,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4381051,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4381143,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4381145,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4381867,I,SI,11,Edwards SAPIEN 3,9600TFX20,2993
4381892,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4382136,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4382175,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4382180,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4382269,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4382341,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4382476,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4382553,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4382570,I,SI,30,Edwards SAPIEN 3,9600TFX26,2993
4383005,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4383090,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4383120,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4383138,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4383138,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4383326,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4383351,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4383351,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4383351,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4383416,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4383416,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4383431,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4383939,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4384150,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4384545,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4384556,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4384641,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4384659,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4384714,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4384835,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4384866,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4384921,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4385170,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4385294,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4385324,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4385343,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4385352,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4385394,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4385767,I,SI,61,Edwards SAPIEN 3,9600TFX26,3190
4385898,I,SI,15,Edwards SAPIEN 3,9600TFX26,2993
4386017,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4386125,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4386125,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4386296,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4386309,I,SI,12,Edwards SAPIEN 3,9600TFX26,2993
4386323,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4386363,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4386363,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4386440,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4386598,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4386598,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4386647,I,SI,5,Edwards SAPIEN 3,9600TFX20,2993
4386914,I,SI,29,Edwards SAPIEN 3,9600TFX29,2993
4387008,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4387057,I,SI,20,Edwards SAPIEN 3,9600TFX26,2993
4387057,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4387057,I,SI,35,Edwards SAPIEN 3,9600TFX26,2993
4387057,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4387170,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4387283,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4387283,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4387341,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4387417,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4387756,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4387839,I,SI,22,Edwards SAPIEN 3,9600TFX26,2993
4387861,I,SI,3,Edwards SAPIEN 3,9600TFX29,3190
4387897,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4387897,I,SI,17,Edwards SAPIEN 3,9600TFX23,3190
4388006,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4388249,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4388330,I,SI,9,Edwards SAPIEN 3,9600TFX29,2993
4388694,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4388808,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4388851,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4388851,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4388947,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4388978,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4389038,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4389063,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4389063,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4389131,I,SI,6,Edwards SAPIEN 3,9600TFX26,3190
4389275,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4389276,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4389375,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4389728,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4389728,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4389744,I,SI,29,Edwards SAPIEN 3,9600TFX26,2993
4389758,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4389758,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4389796,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4389797,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4389797,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4389797,I,SI,3,Edwards SAPIEN 3,9600TFX26,3190
4389813,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4390089,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4390105,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4390215,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4390391,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4390424,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4390707,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4390780,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4390799,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4390823,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4391019,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4391024,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4391311,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4391390,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4391512,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4391512,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4391538,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4391665,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4391677,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4391677,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4391711,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4391711,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4391788,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4391811,I,SI,7,Edwards SAPIEN 3,9600TFX29,2993
4392088,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4392156,I,SI,4,Edwards SAPIEN 3,9600TFX20,2993
4392372,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4392387,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4392390,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4392390,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4392502,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4392667,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4392763,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4392780,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4392807,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4392812,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4392898,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4392919,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4392919,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4392922,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4393000,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4393000,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4393074,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4393083,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4393083,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4393083,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4393102,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4393102,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4393102,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4393140,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4393336,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4393385,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4393421,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4393487,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4393637,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4393641,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4393641,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
4393823,I,SI,9,Edwards SAPIEN 3,9600TFX23,2993
4393987,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4394186,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4394325,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4394325,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4394355,I,SI,11,Edwards SAPIEN 3,9600TFX26,2993
4394407,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4394446,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4394465,I,SI,4,Edwards SAPIEN 3,9600TFX29,3190
4394561,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4394682,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4394682,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4394840,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4394884,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4394982,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4394982,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4394982,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4395040,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4395084,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4395102,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4395102,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4395102,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4395113,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4395131,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4395152,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4395163,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4395239,I,SI,30,Edwards SAPIEN 3,9600TFX26,2993
4395337,I,SI,2,Edwards SAPIEN 3,9600TFX20,2993
4395685,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4395700,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4395875,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4395875,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4395875,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4395875,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4395875,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4395875,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4396135,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4396199,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4396201,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4396282,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4396285,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4396420,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4396459,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4396470,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4396470,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4396470,I,SI,3,Edwards SAPIEN 3,9600TFX29,3190
4396484,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4396564,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4396564,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4397358,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4397363,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4397372,I,SI,7,Edwards SAPIEN 3,9600TFX29,2993
4397493,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4397846,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4398022,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4398047,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4398047,I,SI,13,Edwards SAPIEN 3,9600TFX29,2993
4398187,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4398292,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4398403,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4398608,I,SI,12,Edwards SAPIEN 3,9600TFX23,2993
4398626,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4398626,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4398685,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4398708,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4398953,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4398987,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4399076,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4399373,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4399402,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4399562,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4399562,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4399562,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4399562,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4399593,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4399816,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4400119,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4400263,I,SI,19,Edwards SAPIEN 3,9600TFX23,3190
4400297,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4400303,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4400308,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4400315,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4400389,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4400389,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4400410,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4400410,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4400410,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4400410,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4400481,I,SI,8,Edwards SAPIEN 3,9600TFX29,2993
4400516,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4400535,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4400668,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4400763,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4400816,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4400829,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4401118,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4401165,I,SI,14,Edwards SAPIEN 3,9600TFX23,2993
4401224,I,SI,13,Edwards SAPIEN 3,9600TFX26,3190
4401310,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4401356,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4401356,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4401398,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4401430,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4401557,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4401557,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4401729,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4401732,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4401889,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4402059,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4402127,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4402229,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4402299,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4402306,I,SI,13,Edwards SAPIEN 3,9600TFX29,2993
4402493,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4402493,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4402493,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4402508,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4402622,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4402719,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4402723,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4402950,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4403221,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4403414,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4403414,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4403430,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4403553,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4403583,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4403679,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4403679,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4403705,I,SI,8,Edwards SAPIEN 3,9600TFX29,2993
4403826,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4403826,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4403912,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4404058,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4404329,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4404329,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4404380,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4404440,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4404531,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4404631,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4404661,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4405080,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4405099,I,SI,13,Edwards SAPIEN 3,9600TFX23,2993
4405099,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4405099,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4405099,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4405160,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4405255,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4405255,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4405255,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4405255,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4405310,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4405449,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4405537,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4405578,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4405635,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4405693,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4405923,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4405993,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4406332,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4406409,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4406860,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4406983,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4407215,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4407267,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4407385,I,SI,15,Edwards SAPIEN 3,9600TFX29,3190
4407385,I,SI,9,Edwards SAPIEN 3,9600TFX29,3190
4407429,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4407431,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4407737,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4407866,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4408327,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4408345,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4408398,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4408398,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4408398,I,SI,24,Edwards SAPIEN 3,9600TFX23,3190
4408561,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4408570,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4408587,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4408604,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4408665,I,SI,3,Edwards SAPIEN 3,9600TFX26,3190
4408695,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4408726,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4408759,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4408836,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4409039,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4409187,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4409206,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4409346,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4409346,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4409595,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4409638,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4409666,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4409679,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4409682,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4409798,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4409955,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4409962,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4409975,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4410061,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4410063,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4410114,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4410218,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4410218,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4410326,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4410850,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4411021,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4411240,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4411755,I,SI,3,Edwards SAPIEN 3,9600TFX29,3190
4411767,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4411767,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4411786,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4411958,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4412188,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4412188,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4412286,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4412286,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4412647,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4412788,I,SI,7,Edwards SAPIEN 3,9600TFX23,2993
4412910,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4413469,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4413664,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4413697,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4413750,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4413756,I,SI,10,Edwards SAPIEN 3,9600TFX23,2993
4414042,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4414066,I,SI,5,Edwards SAPIEN 3,9600TFX23,3190
4414066,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4414066,I,SI,3,Edwards SAPIEN 3,9600TFX23,3190
4414131,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4414201,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4414477,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4414589,I,SI,9,Edwards SAPIEN 3,9600TFX23,2993
4414589,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4414655,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4414765,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4414777,I,SI,10,Edwards SAPIEN 3,9600TFX29,3190
4414781,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4414781,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4414799,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4414806,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4415120,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4415196,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4415322,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4415402,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4415404,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4415470,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4415484,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4415524,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4415671,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4415716,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4415747,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4415815,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4415853,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4416136,I,SI,7,Edwards SAPIEN 3,9600TFX29,2993
4416326,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4416425,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4416898,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4417094,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4417276,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4417293,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4417402,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4417490,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4417575,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4417713,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4417837,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4417861,I,SI,39,Edwards SAPIEN 3,9600TFX23,3190
4417962,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4418195,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4418471,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4418471,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4419234,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4419513,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4419517,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4419517,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4419529,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4419629,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4419735,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4419740,I,SI,22,Edwards SAPIEN 3,9600TFX29,2993
4419749,I,SI,8,Edwards SAPIEN 3,9600TFX23,3190
4419789,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4419931,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4419931,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4420261,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4420454,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4420454,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4420591,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4420668,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4420831,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4420922,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4421118,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4421158,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4421158,I,SI,7,Edwards SAPIEN 3,9600TFX26,3190
4421406,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4421630,I,SI,22,Edwards SAPIEN 3,9600TFX29,2993
4421726,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4422277,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4422277,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4422503,I,SI,2,Edwards SAPIEN 3,9600TFX26,3190
4422543,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4422902,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4423081,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4423309,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4423481,I,SI,5,Edwards SAPIEN 3,9600TFX23,3190
4423527,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4423723,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4423723,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4423908,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4423998,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4424076,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4424211,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4424294,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4424398,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4424398,I,SI,27,Edwards SAPIEN 3,9600TFX23,3190
4424525,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4425220,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4425226,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4425233,I,SI,17,Edwards SAPIEN 3,9600TFX26,3190
4425258,I,SI,28,Edwards SAPIEN 3,9600TFX29,3190
4425597,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4425791,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4425838,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4425997,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4426146,I,SI,15,Edwards SAPIEN 3,9600TFX23,2993
4426187,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4426187,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4426187,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4426188,I,SI,3,Edwards SAPIEN 3,9600TFX23,3190
4426301,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4426301,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4426341,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4426341,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4426343,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4426410,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4426432,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4426451,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4426546,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4426577,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4426620,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4426709,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4426709,I,SI,28,Edwards SAPIEN 3,9600TFX20,3190
4426808,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4426817,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4426987,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4427178,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4427553,I,SI,4,Edwards SAPIEN 3,9600TFX20,3190
4427726,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4427836,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4427904,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4427933,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4427993,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4428028,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4428205,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4428719,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4428978,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4428978,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4428978,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4429097,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4429233,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4429256,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4429450,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4429739,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4430087,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4430173,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4430208,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4430251,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4430251,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4430275,I,SI,12,Edwards SAPIEN 3,9600TFX26,3190
4430353,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4430394,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4430463,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4430590,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4431064,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4431111,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4431128,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431128,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4431223,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4431223,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4431223,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4431269,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431269,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431269,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431275,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4431286,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431350,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4431533,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431533,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431533,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4431611,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431686,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4431686,I,SI,7,Edwards SAPIEN 3,9600TFX23,3190
4431688,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4431720,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4431830,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4431961,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4432033,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4432085,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4432095,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4432119,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4432141,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4432354,I,SI,6,Edwards SAPIEN 3,9600TFX29,3190
4432451,I,SI,4,Edwards SAPIEN 3,9600TFX29,3190
4433278,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4433308,I,SI,8,Edwards SAPIEN 3,9600TFX23,3190
4433308,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4433308,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4433308,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4433321,I,SI,2,Edwards SAPIEN 3,9600TFX20,3190
4433330,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4433356,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4433356,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4433407,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4433437,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4433437,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4433457,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4433461,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4433504,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4433504,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4433507,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4433552,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4433823,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4433826,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4433959,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4433988,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4433988,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4433988,I,SI,15,Edwards SAPIEN 3,9600TFX23,3190
4434027,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4434163,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4434206,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4434319,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4434319,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4434579,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4434598,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4434643,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4434906,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4434942,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4435107,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4435149,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4435149,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4435245,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4435444,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4435472,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4435611,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4435611,I,SI,34,Edwards SAPIEN 3,9600TFX29,3190
4435627,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4435938,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4436087,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4436314,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4436575,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4436575,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4436776,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4436804,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4436889,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4437420,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4437420,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4437420,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4437439,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4437452,I,SI,1,Edwards SAPIEN 3,9600TFX20,3190
4437452,I,SI,2,Edwards SAPIEN 3,9600TFX20,3190
4437452,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4437458,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4437458,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4437504,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4437507,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4437546,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4437546,I,SI,9,Edwards SAPIEN 3,9600TFX26,3190
4437760,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4437785,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4437919,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4438132,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4438202,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4438243,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4438394,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4438465,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4438502,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4438502,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4438564,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4438651,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4438690,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4438739,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4438793,I,SI,21,Edwards SAPIEN 3,9600TFX26,2993
4439001,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4439001,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4439008,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4439120,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4439193,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4439257,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4439257,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4439382,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4439548,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4439548,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4439559,I,SI,3,Edwards SAPIEN 3,9600TFX20,2993
4439673,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4439804,I,SI,9,Edwards SAPIEN 3,9600TFX29,2993
4439924,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4440038,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4440087,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4440099,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4440131,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4440220,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4440244,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4440247,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4440247,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4440247,I,SI,34,Edwards SAPIEN 3,9600TFX29,3190
4440278,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4440405,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4440421,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4440517,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4440524,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4440620,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4440816,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4440816,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4440816,I,SI,5,Edwards SAPIEN 3,9600TFX23,3190
4441013,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4441013,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4441082,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4441368,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4441368,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4441368,I,SI,3,Edwards SAPIEN 3,9600TFX23,3190
4441743,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4441781,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4441793,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4441831,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4441903,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4441903,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4441907,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4442133,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4442133,I,SI,8,Edwards SAPIEN 3,9600TFX26,3190
4442159,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4442409,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4442458,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4442560,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4442593,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4442621,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4442726,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4442756,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4442756,I,SI,4,Edwards SAPIEN 3,9600TFX29,3190
4442818,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4442833,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4442846,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4443028,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4443124,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4443124,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4443173,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4443369,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4443455,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4443469,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4443477,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4443557,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4444038,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4444238,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4444245,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4444444,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4444534,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4444589,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4444672,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4444680,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4445116,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4445387,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4445642,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4445664,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4445816,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4445816,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4445848,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4445930,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4445930,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4446059,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4446143,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4446158,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4446206,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4446279,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4446282,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4446318,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4446421,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4446658,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4446727,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4446810,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4446858,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4446988,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4447263,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4447287,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4447346,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4447346,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4447354,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4447382,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4447404,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4447465,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4447498,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4447623,I,SI,7,Edwards SAPIEN 3,9600TFX26,3190
4447740,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4447998,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4448067,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4448149,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4448155,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4448168,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4448427,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4448466,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4448548,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4448605,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4448614,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4448629,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4448629,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4448629,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4448672,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4448708,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4448708,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4449045,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4449194,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4449194,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4449212,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4449291,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4449348,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4449420,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4449558,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4449625,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4449649,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4449672,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4449722,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4449722,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4449851,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4449851,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4449851,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4450169,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4450216,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4450394,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4450598,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4450697,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4450796,I,SI,17,Edwards SAPIEN 3,9600TFX29,3190
4450823,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4450823,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4450823,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4450849,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4450849,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4450849,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4450940,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4450958,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4450958,I,SI,23,Edwards SAPIEN 3,9600TFX20,3190
4451042,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4451085,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4451096,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4451104,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4451131,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4451131,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4451196,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4451304,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4451368,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4451424,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4451519,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4451618,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4451618,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4451620,I,SI,5,Edwards SAPIEN 3,9600TFX26,3190
4451830,I,SI,28,Edwards SAPIEN 3,9600TFX26,3190
4451892,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4451940,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4451970,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4451988,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4451988,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4452095,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4452225,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4452410,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4452593,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4452759,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4452759,I,SI,7,Edwards SAPIEN 3,9600TFX26,2993
4453309,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4453342,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4453342,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4453342,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4453453,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4453564,I,SI,42,Edwards SAPIEN 3,9600TFX26,3190
4453674,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4453737,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4453768,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4453843,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4453987,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4454039,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4454084,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4454167,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4454248,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4454329,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4454477,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4454781,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4454790,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4454861,I,SI,17,Edwards SAPIEN 3,9600TFX26,2993
4454861,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4454879,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4454880,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4455050,I,SI,14,Edwards SAPIEN 3,9600TFX26,2993
4455050,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4455104,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4455104,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4455124,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4455130,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4455501,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4455596,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4455605,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4455694,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4455781,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4456089,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4456152,I,SI,9,Edwards SAPIEN 3,9600TFX20,2993
4456328,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4456534,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4456651,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4456967,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4457066,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4457204,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4457250,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4457472,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4457512,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4457710,I,SI,6,Edwards SAPIEN 3,9600TFX29,3190
4457751,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4458000,I,SI,10,Edwards SAPIEN 3,9600TFX23,2993
4458342,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4458342,I,SI,14,Edwards SAPIEN 3,9600TFX29,2993
4458342,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4458616,I,SI,1,Edwards SAPIEN 3,9600TFX20,3190
4458649,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4458690,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4458800,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4458861,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4458941,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4458941,I,SI,3,Edwards SAPIEN 3,9600TFX26,3190
4459038,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4459156,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4459413,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4459444,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4459612,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4459671,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4459865,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4459886,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4459987,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4460055,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4460067,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4460067,I,SI,36,Edwards SAPIEN 3,9600TFX26,3190
4460067,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4460067,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4460135,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4460174,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4460191,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4460594,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4460758,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4460838,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4461129,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4461201,I,SI,4,Edwards SAPIEN 3,9600TFX26,3190
4461204,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4461254,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4461254,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4461452,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4461452,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4461830,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4461830,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4461830,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4461830,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4461830,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4461833,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4461962,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4462027,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4462221,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4462396,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4462396,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4462675,I,SI,11,Edwards SAPIEN 3,9600TFX23,2993
4462675,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4462734,I,SI,12,Edwards SAPIEN 3,9600TFX23,3190
4462802,I,SI,9,Edwards SAPIEN 3,9600TFX29,2993
4462846,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4462847,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4462938,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4463051,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4463071,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4463109,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4463343,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4463343,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4463445,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4463594,I,SI,18,Edwards SAPIEN 3,9600TFX26,3190
4463760,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4463760,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4463847,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4463876,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4464047,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4464072,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4464115,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4464154,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4464262,I,SI,10,Edwards SAPIEN 3,9600TFX29,2993
4464376,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4464487,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4464745,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4464847,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4464867,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4465427,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4465485,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4465499,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4465541,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4465627,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4465692,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4465720,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4465954,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4466029,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4466082,I,SI,34,Edwards SAPIEN 3,9600TFX26,3190
4466361,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4466517,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4466517,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4466543,I,SI,15,Edwards SAPIEN 3,9600TFX23,2993
4466622,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4466846,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4466885,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4466885,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4466894,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4466903,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4467080,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4467293,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4467345,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4467424,I,SI,4,Edwards SAPIEN 3,9600TFX20,2993
4467826,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4468002,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4468018,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4468184,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4468274,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4468309,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4468315,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4468518,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4468871,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4469178,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4469191,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4469319,I,SI,23,Edwards SAPIEN 3,9600TFX26,2993
4469337,I,SI,34,Edwards SAPIEN 3,9600TFX29,3190
4469535,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4469545,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4469594,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4469638,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4469743,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4469853,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4470145,I,SI,11,Edwards SAPIEN 3,9600TFX29,2993
4470180,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4470202,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4470404,I,SI,8,Edwards SAPIEN 3,9600TFX23,3190
4470557,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4470891,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4470891,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4470891,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4470953,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4470953,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4470953,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4470999,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4471133,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4471235,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4471257,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4471352,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4471501,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4471554,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4471593,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4471701,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4471815,I,SI,7,Edwards SAPIEN 3,9600TFX29,2993
4471837,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4471919,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4471929,I,SI,3,Edwards SAPIEN 3,9600TFX26,3190
4472158,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4472402,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4472431,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4472489,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4472552,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4472859,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4472879,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4472887,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4473044,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4473105,I,SI,32,Edwards SAPIEN 3,9600TFX26,3190
4473141,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4473141,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4473141,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4473311,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4473375,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4473389,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4473442,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4473449,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4473561,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4473588,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4473621,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4473693,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4473881,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4473881,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4474035,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4474045,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4474055,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4474065,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4474078,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4474078,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4474156,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4474297,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4474297,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4474297,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4474297,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4474340,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4474340,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4474436,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4474751,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4474944,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4475090,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4475100,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4475259,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4475373,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4475399,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4475399,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4475565,I,SI,2,Edwards SAPIEN 3,9600TFX29,3190
4475789,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4476032,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4476100,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4476100,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4476103,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4476446,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4476553,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4476631,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4476661,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4476729,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4476804,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4476865,I,SI,2,Edwards SAPIEN 3,9600TFX20,2993
4476967,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4476999,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4477041,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4477052,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4477058,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4477116,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4477141,I,SI,14,Edwards SAPIEN 3,9600TFX23,2993
4477228,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4477257,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4477461,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4477461,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4477479,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4477498,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4477526,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4477526,I,SI,24,Edwards SAPIEN 3,9600TFX29,3190
4477637,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4477696,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4477696,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4477814,I,SI,3,Edwards SAPIEN 3,9600TFX29,3190
4477880,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4477882,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4477907,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4477907,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4477907,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4477907,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4477926,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4477926,I,SI,6,Edwards SAPIEN 3,9600TFX23,3190
4477927,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4477934,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4477969,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4478050,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4478108,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4478129,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4478259,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4478295,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4478320,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4478326,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4478337,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4478356,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4478388,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4478507,I,SI,7,Edwards SAPIEN 3,9600TFX29,2993
4478572,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4478593,I,SI,48,Edwards SAPIEN 3,9600TFX26,3190
4478636,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4478954,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4478954,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4478955,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4479006,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4479188,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4479188,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4479259,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4479329,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4479510,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4479540,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4479727,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4479819,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4479865,I,SI,23,Edwards SAPIEN 3,9600TFX26,2993
4479886,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4479934,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4479999,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4480011,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4480015,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4480015,I,SI,3,Edwards SAPIEN 3,9600TFX23,3190
4480079,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4480079,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4480093,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4480116,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4480356,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4480396,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4480405,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4480405,I,SI,7,Edwards SAPIEN 3,9600TFX23,3190
4480419,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4480428,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4480428,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4480470,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4480490,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4480552,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4480574,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4480665,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4480691,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4480691,I,SI,4,Edwards SAPIEN 3,9600TFX29,3190
4480691,I,SI,16,Edwards SAPIEN 3,9600TFX29,3190
4480848,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4480916,I,SI,-41,Edwards SAPIEN 3,9600TFX23,2993
4480945,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4481012,I,SI,3,Edwards SAPIEN 3,9600TFX20,2993
4481013,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4481021,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4481021,I,SI,14,Edwards SAPIEN 3,9600TFX23,3190
4481050,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4481050,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4481163,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4481170,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4481180,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4481195,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4481198,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4481201,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4481215,I,SI,2,Edwards SAPIEN 3,9600TFX20,2993
4481231,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4481260,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4481260,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4481273,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4481276,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4481358,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4481432,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4481432,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4481630,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4481630,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4481781,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4481894,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4481983,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
4482018,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4482021,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4482131,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4482131,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4482220,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4482223,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4482264,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4482264,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4482298,I,SI,3,Edwards SAPIEN 3,9600TFX23,2993
4482298,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4482298,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4482298,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4482395,I,SI,7,Edwards SAPIEN 3,9600TFX29,2993
4482402,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4482441,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4482468,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4482539,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4482673,I,SI,9,Edwards SAPIEN 3,9600TFX23,2993
4482706,I,SI,18,Edwards SAPIEN 3,9600TFX29,2993
4482723,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4482786,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4482823,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4482823,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4482833,I,SI,2,Edwards SAPIEN 3,9600TFX20,2993
4482971,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4483048,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4483083,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4483144,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4483196,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4483397,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4483477,I,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4483698,I,SI,15,Edwards SAPIEN 3,9600TFX23,2993
4484034,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4484034,I,SI,3,Edwards SAPIEN 3,9600TFX23,3190
4484096,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4484178,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4484278,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4484278,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4484278,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4484305,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4484306,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4484433,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4484487,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4484533,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4484651,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4484659,I,SI,10,Edwards SAPIEN 3,9600TFX23,2993
4484659,I,SI,6,Edwards SAPIEN 3,9600TFX23,3190
4484659,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4484676,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4484758,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
4484791,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4484899,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4484922,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4484998,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4484998,I,SI,13,Edwards SAPIEN 3,9600TFX23,3190
4484998,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4484998,I,SI,13,Edwards SAPIEN 3,9600TFX23,3190
4485002,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4485002,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4485006,I,SI,11,Edwards SAPIEN 3,9600TFX23,2993
4485034,I,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4485047,I,SI,8,Edwards SAPIEN 3,9600TFX23,3190
4485086,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4485151,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4485184,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4485222,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4485256,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4485265,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4485314,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
4485340,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4485530,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4485532,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4485625,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4485721,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4485746,I,SI,14,Edwards SAPIEN 3,9600TFX26,2993
4485780,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4485928,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4485946,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4485969,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4485978,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4486000,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4486143,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4486167,I,SI,5,Edwards SAPIEN 3,9600TFX29,2993
4486190,I,SI,15,Edwards SAPIEN 3,9600TFX23,3190
4486203,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4486552,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4486632,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4486767,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4486877,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4486961,I,SI,16,Edwards SAPIEN 3,9600TFX26,3190
4486993,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4486997,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4487039,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4487275,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4487437,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4487439,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4487480,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4487560,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4487864,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4487887,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4487897,I,SI,11,Edwards SAPIEN 3,9600TFX29,2993
4487897,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4487897,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4487902,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4487998,I,SI,7,Edwards SAPIEN 3,9600TFX23,2993
4488113,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4488113,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4488113,I,SI,19,Edwards SAPIEN 3,9600TFX23,3190
4488133,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4488229,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4488263,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4488295,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4488314,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4488494,I,SI,4,Edwards SAPIEN 3,9600TFX23,3190
4488525,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4488546,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4488614,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4488647,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4488724,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4488736,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4488776,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4488840,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4488868,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4489060,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4489088,I,SI,0,Edwards SAPIEN 3,9600TFX20,2993
4489094,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4489104,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4489184,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4489184,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4489184,I,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4489228,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4489309,I,SI,5,Edwards SAPIEN 3,9600TFX23,2993
4489339,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4489347,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4489356,I,SI,0,Edwards SAPIEN 3,9600TFX20,3190
4489419,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4489596,I,SI,2,Edwards SAPIEN 3,9600TFX23,3190
4489608,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4489625,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4489630,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4489631,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4489656,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4489696,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4489776,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4489778,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4489802,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4489809,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4489811,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4489811,I,SI,14,Edwards SAPIEN 3,9600TFX23,2993
4489811,I,SI,8,Edwards SAPIEN 3,9600TFX23,2993
4489811,I,SI,14,Edwards SAPIEN 3,9600TFX23,3190
4489825,I,SI,8,Edwards SAPIEN 3,9600TFX29,2993
4489866,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4489866,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4489903,I,SI,8,Edwards SAPIEN 3,9600TFX26,2993
4489908,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4489945,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4489952,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4489960,I,SI,2,Edwards SAPIEN 3,9600TFX29,2993
4489962,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4489962,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4490049,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4490172,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4490324,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4490625,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4490633,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4490648,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4490705,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4490791,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4490791,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4490893,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4490893,I,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4490893,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4490893,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4490899,I,SI,2,Edwards SAPIEN 3,9600TFX23,2993
4491010,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4491025,I,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4491025,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4491060,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4491156,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4491286,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4491286,I,SI,13,Edwards SAPIEN 3,9600TFX26,3190
4491286,I,SI,4,Edwards SAPIEN 3,9600TFX26,3190
4491286,I,SI,10,Edwards SAPIEN 3,9600TFX26,3190
4491506,I,SI,9,Edwards SAPIEN 3,9600TFX29,2993
4491657,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4491657,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
4491798,I,SI,6,Edwards SAPIEN 3,9600TFX26,2993
4491868,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4491890,I,SI,1,Edwards SAPIEN 3,9600TFX26,3190
4492003,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4492031,I,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4492126,I,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4492171,I,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4492192,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4492360,I,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4492502,I,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4492617,I,SI,9,Edwards SAPIEN 3,9600TFX26,2993
4492782,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4492908,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4493157,I,SI,2,Edwards SAPIEN 3,9600TFX26,2993
4354195,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
3762722,I,SI,1,Edwards SAPIEN 3,9600TFX29,2993
3762722,I,SI,1,Edwards SAPIEN 3,9600TFX29,3190
4353523,I,SI,6,Edwards SAPIEN 3,9600TFX29,2993
4353523,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4421726,I,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4485256,I,SI,5,Edwards SAPIEN 3,9600TFX26,2993
4081816,I,SI,15,Edwards SAPIEN 3,9600TFX23,2993
4108025,I,SI,134,Edwards SAPIEN 3,9600TFX29,2993
4368391,I,SI,36,Edwards SAPIEN 3,9600TFX26,3190
4375572,I,SI,3,Edwards SAPIEN 3,9600TFX26,2993
4384074,I,SI,4,Edwards SAPIEN 3,9600TFX29,2993
4410419,I,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4417415,I,SI,0,Edwards SAPIEN 3,9600TFX29,3190
4442057,I,SI,46,Edwards SAPIEN 3,9600TFX26,2993
4445059,I,SI,3,Edwards SAPIEN 3,9600TFX29,2993
3864511,S,SI,4,Edwards SAPIEN 3,9600TFX23,2993
3895569,S,SI,437,Edwards SAPIEN 3,9600TFX29,3190
3391171,S,SI,30,Edwards SAPIEN 3,9600TFX23,3190
3391171,S,SI,133,Edwards SAPIEN 3,9600TFX23,3190
3933860,S,SI,368,Edwards SAPIEN 3,9600TFX29,2993
3956941,S,SI,66,Edwards SAPIEN 3,9600TFX29,3190
3975567,S,SI,85,Edwards SAPIEN 3,9600TFX29,3190
3977126,S,SI,123,Edwards SAPIEN 3,9600TFX23,3190
4029079,S,SI,331,Edwards SAPIEN 3,9600TFX26,3190
4035735,S,SI,32,Edwards SAPIEN 3,9600TFX23,3190
4037947,S,SI,186,Edwards SAPIEN 3,9600TFX26,3190
3969095,S,SI,42,Edwards SAPIEN 3,9600TFX23,3190
4022473,S,SI,30,Edwards SAPIEN 3,9600TFX26,2993
4079767,S,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4087218,S,SI,125,Edwards SAPIEN 3,9600TFX29,2993
4102651,S,SI,135,Edwards SAPIEN 3,9600TFX29,3190
4119453,S,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4127077,S,SI,187,Edwards SAPIEN 3,9600TFX26,3190
4042643,S,SI,25,Edwards SAPIEN 3,9600TFX26,2993
4048573,S,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4150384,S,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4160452,S,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4174871,S,SI,4,Edwards SAPIEN 3,9600TFX26,2993
4203217,S,SI,9,Edwards SAPIEN 3,9600TFX23,2993
4237840,S,SI,1,Edwards SAPIEN 3,9600TFX20,2993
4240483,S,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4245706,S,SI,343,Edwards SAPIEN 3,9600TFX23,3190
4252198,S,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4253569,S,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4257594,S,SI,0,Edwards SAPIEN 3,9600TFX26,2993
4260521,S,SI,6,Edwards SAPIEN 3,9600TFX20,2993
4268017,S,SI,23,Edwards SAPIEN 3,9600TFX26,3190
4268017,S,SI,48,Edwards SAPIEN 3,9600TFX26,3190
4275969,S,SI,1,Edwards SAPIEN 3,9600TFX23,2993
4287444,S,SI,4,Edwards SAPIEN 3,9600TFX23,2993
4293727,S,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4295027,S,SI,1,Edwards SAPIEN 3,9600TFX23,3190
4308693,S,SI,12,Edwards SAPIEN 3,9600TFX23,3190
4331566,S,SI,22,Edwards SAPIEN 3,9600TFX29,3190
4337693,S,SI,0,Edwards SAPIEN 3,9600TFX29,2993
4349343,S,SI,1,Edwards SAPIEN 3,9600TFX26,2993
4351861,S,SI,0,Edwards SAPIEN 3,9600TFX26,3190
4354725,S,SI,0,Edwards SAPIEN 3,9600TFX23,2993
4357055,S,SI,6,Edwards SAPIEN 3,9600TFX23,2993
4357230,S,SI,0,Edwards SAPIEN 3,9600TFX23,3190
4363242,S,SI,15,Edwards SAPIEN 3,9600TFX26,2993
4373962,S,SI,5,Edwards SAPIEN 3,9600TFX26,3190
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEATH VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
frances preston
1 edwards way
irvine, CA 92614
9492505190
MDR Report Key7245033
MDR Text Key99154501
Report Number2015691-2018-00378
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Model Number9600TFX
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-