• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913); Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/17/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A medtronic representative inspected the imaging system onsite and confirmed the issue.The door was slightly open and the rotor was in the wrong position.The door was repaired and the issue resolved.A full imaging system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.No parts were replaced or returned to the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while in a spinal fusion procedure the imaging system would not open.Manual override was attempted but there was a 'clicking noise' and the door would not fully open.The system was rebooted and then stated, "initializing, please wait".E-stop was on the pendant and it was not possible to turn it off.The site attempted to manually open the door, but it was noted that at a point the wrench would no longer move and they could only see orange.The site again rebooted, attempted to open and do an override.The system would not open.The patient was removed from the imaging system during trouble shooting and the case was completed with a new imaging system.There was no impact on the patient outcome as a result of this issue.No additional information was provided.
 
Manufacturer Narrative
Additional information: additional procedure information received.
 
Event Description
Medtronic received information that there was at least a 30 minute delay to the procedure.
 
Manufacturer Narrative
Patient identifying number could not be obtained from the site.Patient information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7248933
MDR Text Key99269729
Report Number1723170-2018-00586
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/17/2018
Initial Date FDA Received02/07/2018
Supplement Dates Manufacturer Received05/03/2018
09/07/2018
Supplement Dates FDA Received06/01/2018
09/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age73 YR
Patient Weight83
-
-