MEDTRONIC EUROPE SARL VIVA QUAD XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTBA2Q1 |
Device Problems
Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
|
Patient Problem
Shock from Patient Lead(s) (3162)
|
Event Date 12/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: 4298 lead, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient received a suspected inappropriate shock from the cardiac resynchronization therapy defibrillator (crt-d) for atrial flutter with rapid ventricular response.It was noted the initial detection was withheld for atrial flutter, however, the episode progressed for 49 minutes and reached a dual tach detection ventricular tachycardia (vt) with supra ventricular tachycardia (svt), which the device delivered therapy for the dual tachycardia.It was noted the therapy "broke" the arrhythmia and the device remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received, which indicated a new wavelet was taken as a result of the event.No patient complications have been reported as a result of this event.
|
|