• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL VIVA QUAD XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL VIVA QUAD XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA2Q1
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 12/09/2017
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 4298 lead, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient received a suspected inappropriate shock from the cardiac resynchronization therapy defibrillator (crt-d) for atrial flutter with rapid ventricular response.It was noted the initial detection was withheld for atrial flutter, however, the episode progressed for 49 minutes and reached a dual tach detection ventricular tachycardia (vt) with supra ventricular tachycardia (svt), which the device delivered therapy for the dual tachycardia.It was noted the therapy "broke" the arrhythmia and the device remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received, which indicated a new wavelet was taken as a result of the event.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA QUAD XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7255917
MDR Text Key99842270
Report Number9614453-2018-00272
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2019
Device Model NumberDTBA2Q1
Device Catalogue NumberDTBA2Q1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/12/2017
Initial Date FDA Received02/08/2018
Supplement Dates Manufacturer Received02/28/2018
Supplement Dates FDA Received04/10/2018
Date Device Manufactured09/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
-
-