Model Number DTBA2D1
Device Problems Defibrillation/Stimulation Problem (1573); Invalid Sensing (2293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2017
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 6944 lead, implanted: (b)(6) 2004; 4193 lead, implanted: (b)(6)2004.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported by the physician that the cardiac resynchronization therapy defibrillator (crt-d) identified two episodes of suspected ventricular tachycardia (vt) as supra ventricular tachycardia (svt)/atrial fibrillation (af).The physician questioned the detection of the crt-d and the device's decision to withhold therapy for episodes they felt should have treated.The physician noted they wanted the episodes treated the same and reprogramming was discussed.The crt-d remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7257268
MDR Text Key99917331
Report Number9614453-2018-00489
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2018
Device Model NumberDTBA2D1
Device Catalogue NumberDTBA2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2017
Initial Date FDA Received02/08/2018
Date Device Manufactured11/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-