AV-TEMECULA-CT ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Angina (1710); Thrombosis (2100)
Event Date 01/14/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The scaffold remains in the patient.Investigation is not yet complete.A follow up report will be submitted with all relevant information.The absorb device is currently not commercially available in the u.S; however, it is similar to a device sold in the u.S.
 
Event Description
It was reported that the procedure was to treat a lesion in the right coronary artery.The lesion was prepared with a cutting balloon prior to implanting an absorb scaffold.Post-dilatation was performed.The patient returned to the hospital because of angina symptoms at which time thrombosis was observed.Balloon angioplasty was done and a drug eluting stent was implanted for treatment.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Udi#: in the absence of a reported part number, the udi cannot be calculated.There was no reported device malfunction and the product was not returned as the scaffold remains in the anatomy.A review of the lot history record and complaint history of the reported device could not be conducted because the part and lot numbers were not provided.The reported patient effects of angina and thrombosis, as listed in the bioresorbable vascular scaffold (bvs) system, absorb, instructions for use, are known adverse events associated with the use of a coronary scaffold in native coronary arteries.Based on the case information, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Type of Device
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7259570
MDR Text Key99582892
Report Number2024168-2018-00951
Device Sequence Number1
Product Code PNY
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P150023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2018
Initial Date FDA Received02/09/2018
Supplement Dates Manufacturer Received02/16/2018
Supplement Dates FDA Received02/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-