Model Number 816300
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/19/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Per the manufacturer's subsidiary, the unit was being reset for the setup of the date when the issue occurred.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) was black and would not activate.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, product surveillance technician (pst) observed the central control monitor (ccm) to power on with a black screen and a "disk boot failure, instert system disk and press enter" error messaged displayed.The unit was rebooted five times, each resulting with the same error message.The complimentary metal-oxide semiconductor (cmos) battery had a volt reading of.2 (normal voltage is three volts).Information was received from the manufacturer's subsidiary that the batteries and hard drive of the ccm have never been replaced.It is recommended in the service procedure that the batteries and hard drive are replaced every six years.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7263558
MDR Text Key100127494
Report Number1828100-2018-00070
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Initial Date Manufacturer Received 01/31/2018
Initial Date FDA Received02/12/2018
Supplement Dates Manufacturer Received02/15/2018
04/17/2018
Supplement Dates FDA Received03/12/2018
04/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-