ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR, PRODUCT CODE: DRM Back to Search Results
Model Number MODEL 100
Device Problem Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2018
Event Type  malfunction  
Manufacturer Narrative
Zoll has not yet received the autopulse platform for evaluation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
 
Event Description
During patient use, the autopulse platform inserted with an autopulse li-ion battery (sn (b)(4)) performed continuous compression without issue for an unspecified amount of time.The user stopped the platform to exchange to a new battery.Manual cpr was performed by another crew while performing the battery exchange.During the battery exchange, the user was unable to remove the battery out of the platform.The reporter noted physical damage observed on the battery finger latch.The patient continued to receive manual cpr.No known impact or patient consequence was reported.This references the report of the platform unable to power on.The report of physical damage on the battery during use is referenced under mfr 3010617000-2018-00142.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR, PRODUCT CODE: DRM
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7263802
MDR Text Key100131521
Report Number3010617000-2018-00143
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-