• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SHERLOCK 3CG* TCS SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SHERLOCK 3CG* TCS SENSOR Back to Search Results
Catalog Number 9770131
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
The magnet tracking on the sherlock 3cg sensor is not functioning at all.Nurse also reported that she has ordered multiple cxr for procedures due to not trusting the sherlock.Unit has been having issues with ekg as well.During one procedure, nurse was unable to obtain a negative deflection, she ordered a cxr and the picc was in the mid to lower atrium.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SHERLOCK 3CG* TCS SENSOR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key7264327
MDR Text Key99985406
Report Number3006260740-2018-00187
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741124341
UDI-Public(01)00801741124341
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770131
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-