ST. JUDE MEDICAL, INC. MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The result of the analysis confirmed the reported event of failure to start-up.Inspection of the ac power adapter revealed burn marks on the main pcb consistent with damage due to the high current flow through the ac wall power outlet.
 
Event Description
This report is to advise of an event observed during analysis.Burn marks to the main pcb were revealed during analysis.The transmitter was initially returned due to a start-up failure.There were no reported adverse patient consequences related to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7264353
MDR Text Key99900534
Report Number2017865-2018-01277
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2017
Initial Date FDA Received02/12/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-