(b)(4).The expedium straight thoracic pedicle probe was returned to the customer quality unit (cqu).The probe was fractured approximately 40mm from the distal tip.The fractured tip was not returned for evaluation.An optical image of the fractured surface the probe reveals a rough grainy appearance consistent across the entirety of the surface with no evidence of fatigue striations.This suggests the probe underwent a quasi-static overload failure.No material defects or other abnormalities were observed in this analysis.A supplemental hardness test was performed on the driver.This probe is within of the specified guidelines.No issues were identified during the manufacturing and release of this instrument that could contribute to the problem reported by the customer.This instrument was released accomplishing all quality requirements.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the probe tip breaking cannot be determined from the sample and the information provided.The results of fracture analysis have determined that a probable root cause is quasi-static overload failure due to unexpectedly high forces placed on the tip during use.No issues could be identified in the manufacturing or release of this product.Therefore, this complaint will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|