Model Number 173016
Device Problems Detachment Of Device Component (1104); Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/31/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic nissen procedure, the device¿s needle popped out of jaws while being used in wrap.Surgeon was midway through suturing and completed tie with graspers.The needle fell into the patient cavity and was retrieved by graspers.There was no injury caused to the patient and no medical intervention was required.
 
Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.Visual functional indicated no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic nissen procedure, the device¿s needle popped out of jaws while being used in wrap.Surgeon was midway through suturing and completed tie with graspers.The needle fell into the patient cavity and was retrieved by graspers.New device was used successfully with two subsequent loads without issue.There was no injury caused to the patient and no medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO STITCH
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7267234
MDR Text Key99938736
Report Number9612501-2018-00199
Device Sequence Number1
Product Code OCW
UDI-Device Identifier10884521100008
UDI-Public10884521100008
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number173016
Device Catalogue Number173016
Device Lot NumberJ7G1696EX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
Patient Weight91
-
-