STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
|
Back to Search Results |
|
Catalog Number 0250080617 |
Device Problem
Device Packaging Compromised (2916)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: bent shaft, burn on insulation at shaft.Confirmed failure: bent shaft, burn on insulation at shaft.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.(b)(4).
|
|
Event Description
|
It was reported that the insulation had been compromised.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: bent shaft, burn on insulation at shaft.Confirmed failure: bent shaft, burn on insulation at shaft.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.Gtin: (b)(4).
|
|
Event Description
|
It was reported that the insulation had been compromised.
|
|