• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Catalog Number 295025-001
Device Problems Device Displays Incorrect Message (2591); Battery Problem (2885); Charging Problem (2892)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.In addition, patients are provided with several onboard batteries, and the freedom driver has a redundant power source of external wall power.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom onboard battery would charge while in the battery charger, but the patient's freedom driver exhibited an alarm when the battery was inserted into the driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
Investigation testing determined that the freedom onboard battery was over-discharged and had been allowed to remain in that state for an extended period of time to the point where the internal safety monitor initiated its recovery-charge function.With a battery in this state, the freedom driver will annunciate a battery error alarm as reported by the customer.The battery was recovered from the over-discharge condition and successfully passed all sections of the evaluation procedure.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7267971
MDR Text Key100266034
Report Number3003761017-2018-00057
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age52 YR
-
-