The recipient reportedly experienced a skin flap infection.A second degree burn was observed over the implant site.On (b)(6) 2017, the recipient was prescribed mupirocin.The recipient was recommended device non-use for 2 months.Due diligence attempts to obtain additional information regarding recipient status were unsuccessful.The device passed the external visual inspection.The photographic imaging inspection revealed a broken electrode wire near the fantail.This is not believed to be related to the return reason.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.This is the final report.
|