Two devices were received for evaluation.During visual examination, the pouch was observed open; therefore, the peel and seal width test was performed to evaluate the quality of the seal.The results were satisfactory due to the sealed is uniformed across the sealed area.Also, it was observed that the packaging material is correctly aligned.The reported condition was verified; however, the cause of the condition could not be determined.There are controls in place to detect and prevent an opening packaging in the manufacturing.The third device was not returned for evaluation; therefore, a device analysis could not be completed.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|