Model Number 595000-001 |
Device Problems
No Display/Image (1183); Electrical /Electronic Property Problem (1198)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because although the freedom driver display did not show any information, the freedom driver continued to perform its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the freedom driver display did not show any information while supporting the patient.The customer also reported that the patient was switched to a back-up freedom driver.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The customer-reported unresponsive lcd display was confirmed during incoming inspection.During investigation testing, it was determined that the root case was a malfunction of the lcd printed circuit board (pcb).The root cause of the pcb malfunction was the failure of the actuator button associated with powering on the circuit.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|