The device was received for evaluation and the evaluation is complete.Hi-pot and wet functional tests were performed with no issues or errors and all tests passed.A visual inspection of the device was performed using an unaided naked eye.During the visual inspection, it was found that the pump cover was loose and that loctite had not been applied to the shoulder bolt.The reported issue was verified.The cause of the condition was determined to the absence of loctite during the assembly process of manufacturing.To correct the condition, the shoulder bolt would be replaced, and loctite applied.Should additional relevant information become available, a supplemental report will be submitted.
|