• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE 2000 ANTI-TG AB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED IMMULITE 2000 ANTI-TG AB Back to Search Results
Model Number IMMULITE 2000 ANTI-TG AB
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/24/2018
Event Type  malfunction  
Manufacturer Narrative
The customer contacted siemens to report immulite 2000 xpi anti-tg patient results that are in the normal range compared to an abnormal high patient result obtained with anti-tg on the alternate platform which aligns with the patient's clinical condition.The siemens regional support center (rsc) and headquarter support center (hsc) are investigating this issue.
 
Event Description
The customer called siemens to inform that an anti-thyroglobulin (anti-tg) patient result obtained on an alternate platform was abnormal high for anti tg and used to compare against the immulite 2000 xpi anti-tg result.The customer stated that the initial patient test result obtained on the immulite 2000 xpi system is in a normal range and reported to the physician.The initial result was questioned by the physician and the same patient sample was used for repeat testing on an alternate immulite 2000 xpi system.The customer informed that the repeated test result was also in a normal range and not an expected result.The customer stated that the initial result obtained from the alternate platform is correct due to the clinical condition of the patient.There are no reports of patient intervention or adverse health consequence due to the anti-tg discordant patient results.
 
Manufacturer Narrative
Siemens filed the initial mdr on 21-february-2018.Additional information (03-april-2018): a headquarter support center (hsc) specialist has performed an investigation and determined that without having the sample to test, hsc cannot confirm the customer's results or test on an alternate platform.The most likely cause of the difference in results is due differences in preparation and fixation of the antigen used, buffer and technical modalities implemented and diversity of anti-tg antibodies present in the test sample.Hsc has assisted the regional support center (rsc) specialist with references (e.G., articles) from the national center for biotechnology information to help explain the difference in results between the alternate platform and immulite anti-tg ab assays.The rsc specialist has provided the information to the customer and informs that the customer is satisfied and no further assistance or evaluation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMMULITE 2000 ANTI-TG AB
Type of Device
IMMULITE 2000 ANTI-TG AB
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
glyn rhonwy
llanberis, gwynedd LL55 4EL
UK  LL55 4EL
MDR Report Key7287138
MDR Text Key100793151
Report Number2432235-2018-00085
Device Sequence Number1
Product Code JZO
UDI-Device Identifier00630414962108
UDI-Public00630414962108
Combination Product (y/n)N
PMA/PMN Number
K991094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIMMULITE 2000 ANTI-TG AB
Device Lot Number673
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/26/2018
Initial Date FDA Received02/21/2018
Supplement Dates Manufacturer Received04/03/2018
Supplement Dates FDA Received04/25/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age36 YR
-
-