Model Number 3286
Device Problems Disconnection (1171); High impedance (1291); Low impedance (2285)
Patient Problem Inadequate Pain Relief (2388)
Event Date 01/08/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 3: reference mfr.Report#: 1627487-2018-01509, reference mfr.Report#: 1627487-2018-01701.It was reported the patient¿s leads were showing multiple high and low impedances.The patient underwent surgical intervention where one of the leads was replaced with a new one.The physician stated when the pocket incision was opened it was observed that neither of the lead tails were inserted properly in the ipg header.It is unknown which lead was replaced therefore we are reporting on both.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE S-8 LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7287452
MDR Text Key100640565
Report Number1627487-2018-01508
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model Number3286
Device Lot Number5003640
Other Device ID Number05414734402231
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age67 YR
-
-