• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL MIDWEST TRADITION PLUS FIXED BACK END WITH FIBER OPTICS; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL MIDWEST TRADITION PLUS FIXED BACK END WITH FIBER OPTICS; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 770045
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Though no medical/surgical intervention was required to preclude a serious injury in this event, there have been previously reported events involving a similar device that resulted in the need for medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event a doctor reported that a midwest tradition plus handpiece overheated.There was no injury or intervention.
 
Manufacturer Narrative
Evaluation found excessive use/abuse of the cartridge.The turbine was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST TRADITION PLUS FIXED BACK END WITH FIBER OPTICS
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7292847
MDR Text Key101006269
Report Number1419322-2018-00028
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K963050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number770045
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/24/2018
Initial Date FDA Received02/23/2018
Supplement Dates Manufacturer Received04/04/2018
Supplement Dates FDA Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-