• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO CAREASSIST ES; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO CAREASSIST ES; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P1170G0000454
Device Problems Device Operates Differently Than Expected (2913); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/02/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the hand pendant on the bed needed to be replaced.Per the hill-rom service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Patient pendant: disconnect the patient pendant and check the condition of the connector.Then reconnect or replace the pendant.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.Replace the pendant if necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the hand pendant to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the bed's head section goes up automatically by itself and will not go down when pressing the head down button.The bed was located at the account.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREASSIST ES
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key7297519
MDR Text Key101131633
Report Number3006697241-2018-00018
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1170G0000454
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/05/2018
Initial Date FDA Received02/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-