ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3228
Device Problems Disconnection (1171); High impedance (1291); Low impedance (2285)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr report #: 3006705815-2018-00364.It was reported that the patient experienced ineffective therapy.Reprogramming was unable to provide resolution.The patient's lead tested for high and low impedances.Reportedly, x-rays confirmed the lead has pulled out from the ipg.Follow up revealed that the physician was unable to relocate the patient's lead on (b)(6) 2018 due to scar tissue.As a result, surgical intervention may be taken to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7298553
MDR Text Key100998584
Report Number1627487-2018-01670
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2013
Device Model Number3228
Device Lot Number3536587
Other Device ID Number05414734401913
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/30/2018
Initial Date FDA Received02/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL: 1194, SCS ANCHOR (2)
Patient Outcome(s) Other;
Patient Age75 YR
-
-