Model Number CD3231-40 |
Device Problems
Over-Sensing (1438); Device Operates Differently Than Expected (2913)
|
Patient Problems
Complete Heart Block (2627); No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/30/2018 |
Event Type
Injury
|
Event Description
|
It was reported that a patient presented with the device and right ventricular lead oversensing via merlin.Net transmission.Episodes of ams due to oversensing were also noted.Review of the episodes confirmed unspecified oversensing, possibly caused by environmental noise, far-field signals or lead noise.The patient is currently stable and scheduled for revision.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Additional information: noise was reproducible in clinic with isometrics.New information received indicates that the patient presented for a lead revision.Psa testing was clear and noise was not reproducible.Rv lead was capped and replaced.There were a couple episode of noise on competitor's ra lead.The patient became pacemaker dependent possibly due to the rv lead placement on the septum that knocked out the right bundle branch.The patient developed complete heart block by the end of the case.The patient was stable post procedure.
|
|
Manufacturer Narrative
|
No conclusion code available.The reported field event of oversensing was confirmed in the device image.Bench testing found the device to be normal.The cause of the reported event remains undetermined.
|
|
Event Description
|
The device was explanted on (b)(6)2018.
|
|
Search Alerts/Recalls
|